In Compliance with IEC/CEI EN 60529 for IP5X- IP6X protection degrees
Main Features
- Internal Chamber Dimensions 1000x1000x1000 mm WxDxH
- Door with Gaskets and Inspection Window
- Lower collecting hopper
- Compartment for circulation pump
- Vibrating Hopper system to avoid talc deposits
- Talc circulation system by means of radial fan
- Aspiration from the lower part with introduction in the upper part
- Deflector of powder distribution
- Internal socket outlet for specimen connection
- Pre-heating system for internal moisture removal
- Internal/external pressure compensator filter
- Internal upper tracks complete with sliding hooks
- General Switch
- Programmer for cycle timer 24h
- Test cycle selector “Manual –Automatic”
- Selector for internal socket
- Switch for intermittence electrical outlet
- Control switch for fans
- Selector for heating insertion
- Signalling and control lights
- Power supply 400V 3P+N 50hz
- Installed power 1 kW
- Noise exceeding 80 decibels, with vibrator inserted
- External dimensions approx 1160x1080x1960mm WxDxH
Standard Equipment
- 25 kg of talc HM4
- Instruction Manual
- Talc particle size certificate
- Warranty Certificate
- CE Certificate of Conformity
Vacuum-group for the specimen
- Instrument holder structure in powder-coated steel
- Vacuometer for specimen internal vacuum measurements – double scale 40 mbar
- Additional vacuometer with calibration certificate issued by the manufacturer
- Flowmeter for checking the air flowed from the specimen having the following characteristics:
- Minimum flow rate: 40 Lt/h
- Maximum flow rate: 10 mc/h
- Maximum working pressure: 0,5 Bar
- Calibration certificate issued by the manufacturer
- Filter for instrument protection
- Vacuum pump with inverter for the regulation of the vacuum
- Digital safety pressure switch in case of over-depression
Dust Sieve
- Square mesh with diameter of 200 mm
- Nominal distance between wires 5 μm
IP5K and IP6K Testing System
- 6-second air/dust movement with pause every 15 minutes
- Adjustable by PLC
- In compliance with ISO/SIN 40050-1 and SAE J 575